Set-up errors of cone-beam CT in precision radiotherapy for patients with lung cancer
-
GONG Han-shun,
-
WANG Yun-lai,
-
JU Zhong-jian,
-
DU Lei,
-
WANG Jing,
-
XU Shou-ping,
-
OU Guang-ming,
-
GANG Ying,
-
YANG Yan,
-
SUN Bo,
-
ZHOU Xue-hai,
-
HUA Peng
-
-
Abstract
Objective To study the set-up margin of clinical target area by analyzing the set-up errors of cone-beam CT in 16 lung cancer patients undergoing fractioned radiotherapy. Methods Sixteen patients with lung cancer(8 with left lung cancer,5 with right lung cancer and 2 with mediastinal lymphanode metastasis) who underwent image-guided radiation therapy(IGRT)were included in this study.All patients were fixed with thermoplast phantom and underwent IGRT every day.Set-up errors of cone-beam CT were calculated according to its matched and planned CT images in left-right(x),superior-inferior(y) and anterior-posterior(z) directions. Results The set-up error of cone-beam CT in the x,y and z axes was(0.11±0.06)cm,(0.14±0.10)cm and(-0.04±0.06)cm respectively,while the set-up error of cone-beam CT in its rotary axis was 0.57°±0.17°,0.05°±0.20°and-0.01°±0.33°,respectively.The set-up error of cone-beam CT in the y direction was relatively larger than in the x and z direction. Conclusion Application of one-beam CT can significantly reduce the set-up errors of radiotherapy in patients with lung cancer and the effect on PTV margin with an outward extension of 0.62cm,0.82cm and 0.49cm in the x,y and z directions,respectively.
-
-