Application of CT attenuation correction in 99mTc-GSA SPECT liver imaging
-
-
Abstract
Objective To investigate the effect of CT attenuation correction on the 99mTc-GSA SPECT imaging in patients with hepatocellular carcinoma. Methods We selected 24 patients who underwent 99mTc-GSA SPECT/CT scans in the department of nuclear medicine in the first medical center of Chinese PLA General Hospital from March 2017 to August 2018,and calculated the K and FV values with or without CT attenuation correction.The K and FV values were recorded as GSA-KNAC and GSA-FVNAC without CT attenuation correction,and GSA-KAC and GSA-FVAC after CT attenuation correction.The difference was compared between these two groups and their correlations with ICG,and liver serum biochemical indicators were explored. Results The values of GSA-KNAC and GSA-KAC were 0.21±0.06 and 0.20±0.06 respectively,without significant difference (P=0.750).GSA-FVNAC was significantly higher than GSA-FVAC (10.00±2.8 vs 12.64±3.31,P<0.001).GSA-KAC was correlated with INR (r=-0.49),PLT (r=0.49),PTA (r=0.62),PT (r=-0.51),ICG-R15 (r=-0.66),ICG-K (r=0.81) (all P<0.05),and it had the highest correlation with ICG-K.There was no association between GSA-FVNAC and serum biochemical indicators or ICG (P>0.05).GSA-FVAC was correlated with T-BIL and ICG-R15 (r=-0.58 and-0.53,P<0.05,respectively). Conclusion The values of GSA-K and GSA-FV obtained by CT attenuation correction are accurate and can indicate the liver function via 99mTc-SPECT and CT fusion imaging.99mTc-GSA SPECT/CT imaging can facilitate the evaluation of liver function,and it has close correlation with other clinical liver function indicators.
-
-