Evaluation of the Methods for Disinfection of Fibergastroscope

  • 44 samples of fibergastroscope were treated with gastroscope auto-disinfec machine plus oxychlorocide-1000 (a disinfectant). It was found that by bacteriological examination about 99.0—99.9% of contaminating bacteria on the surface of gasreoscope coule be cleared up wh- ile over 99.9% of contaminating bacteria coule be removed with three-step method. It was significantly different by statistical analysis (P<0.05). It is proposed that the machine method used now for disinfection of gastroscope should be further improved.