ZHANG Chuang, QIN Chong, YUAN Xiao-dong, QUAN Chang-bin, QIAO Yuan-gang, AO Guo-kun, TANG Fa-kuan, DAN Zhao-liang. Evaluation of left ventricular venous system with 320-slice dynamic volume computed tomography[J]. ACADEMIC JOURNAL OF CHINESE PLA MEDICAL SCHOOL, 2012, 33(6): 607-609.
Citation: ZHANG Chuang, QIN Chong, YUAN Xiao-dong, QUAN Chang-bin, QIAO Yuan-gang, AO Guo-kun, TANG Fa-kuan, DAN Zhao-liang. Evaluation of left ventricular venous system with 320-slice dynamic volume computed tomography[J]. ACADEMIC JOURNAL OF CHINESE PLA MEDICAL SCHOOL, 2012, 33(6): 607-609.

Evaluation of left ventricular venous system with 320-slice dynamic volume computed tomography

  • Objective To facilitate the implantation of left ventricular(LV) lead in cardiac resynchronization therapy(CRT) by evaluating the LV venous system with 320-slice dynamic volume computed tomography(CT). Methods The LV venous system was reconstructed in 110 patients with coronary heart disease(40 males and 70 females) at the age of 57.9±10.8 years during which the R–R interval of body surface ECG was 30%-80%. Results The LV venous system was clearly observed in all patients.The best imaging was achieved in 91.2% of the patients when the R–R interval of body surface ECG was 30%-50%.One of the 2 most frequently used target veins(postero-lateral vein and lateral vein) in CRT was not found in 23.6%-26.4% of the patients,while both of them were not be observed in 5.5% of the patients. Conclusion 320-slice dynamic volume CT can clearly show the LV venous system and evaluate the most frequently used target veins in CRT,thus contributing to the implantation of LV lead during CRT.