Bioluminescence assay of creatine kinase in serum and It’s Clinical initial application

  • This paper presented an assay method for creatine kinase in serum with bioluminenscence and It’s clinical application. This assay method requied a least amount of reagents and sample size, was easy to perform, and also specific and reproducible. The coefficient of variation within assay is 3.8% and 8.7% between assay rate of recovery, 98.6%. The activities determined were 144.22±52.40, in healthy individual, 1018.39±550.29 in cardiac infarction, and 1661.98±689.27 in myopathy μ mol/L/min. This new method may be of value in the clinicap diagnosis and monitoring of certatn diseases.