医用高能电子直线加速器中子污染防护计算

Thecalculation of protection of neutron contamination from high energy medical electron accelerators

  • Abstract: This paper presents the result of calculation of the dose rate of neutron ontamination rom medical electron accelerators by using Kersey’s method, and giving a practical example. The esult obtained is in coincidence with that of measurement. This work is seems to be very mportant for the design of protection gate of neutron.

     

/